1. No part of this manual may be reproduced in any form without the expressed written permission of SANYO. 2. The contents of this manual are subject to change without notice. 3. Please contact SANYO if any point in this manual is unclear or if there are any inaccuracies. SANYO Electric Biomedical Co., Ltd.
PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person.
Page 4
PRECAUTIONS FOR SAFE OPERATION WARNING Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain water. Only qualified engineers or service personnel should install the unit. The installation by unqualified personnel may cause electric shock or fire. Be sure to install the unit on a sturdy floor.
Page 5
PRECAUTIONS FOR SAFE OPERATION WARNING Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electric shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance.
Page 6
PRECAUTIONS FOR SAFE OPERATION CAUTION Never damage or break the power supply plug or cord. Do not use the supply plug if its cord is loose. This may cause fire or electric shock. Do not touch any electrical parts such as the power supply plug or any switches with a wet hand. This may cause electric shock.
CAUTIONS FOR USAGE 1. If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes. This protects the compressor. 2. This inner cabinet is refrigerated by the forced circulation of cooled air inside the chamber. Ensure that the intake and exhaust vents are not blocked.
CAUTIONS FOR USAGE Model with CE mark This unit is equipped with a circuit breaker on the back. Make sure this breaker is switched ON before the unit starts to run. When the operation of the unit is stopped by this breaker, contact with dealer or service station after disconnected the power supply plug.
NAME AND FUNCTION OF PARTS Refrigerator Freezer 1. Glass window: Water can sometimes condense on the glass in areas of high humidity. Wipe off the condensation with a dry soft cloth. 2. Magnetic door gasket: Always keep clean. 3. Control panel: The panel is divided into two portion; Upper portion is an indicator and lower is a controller.
NAME AND FUNCTION OF PARTS Control panel 1. Alarm buzzer stop key (BZ): Press this key to silence the buzzer in the event that the alarm operates and buzzer sounds. 2. Display select key (REF./FREEZ.): By pressing this key, internal temperature display of refrigerator or freezer and each setting is selected.
BEFORE COMMENCING OPERATION Installation site To operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions: 1. A location not subjected to direct sunlight Installation in a location subjected to direct sunlight may lead to inadequate cooling. 2.
BEFORE COMMENCING OPERATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a neutral detergent and wipe it up with a wet cloth. 2.
OPERATING INSTRUCTIONS Control panel 1. Setup of chamber temperature Table 1 shows the basic operation method. Perform key operation in the sequence indicated in the table. The example in the table is based on the assumption that the refrigerator temperature is 4 C and the freezer temperature is -25 Note:...
Page 14
OPERATING INSTRUCTIONS 2. Key lock operation This unit incorporates a key lock feature which can inhibit the tampering using the keys on the control panel. The key lock is set to OFF at the factory. Display Mode Function Key lock OFF Temperature change enabled Key lock ON Temperature change disabled...
Page 15
OPERATING INSTRUCTIONS Alarm and safety functions This unit has the alarm and safety functions shown in Table 3, and also a self diagnostic function. Kind of alarm Situation Indication Buzzer Safety operation or safety High Refrigerator Alarm lamp flashes. Intermittent tone Remote alarm is temperature If the chamber temperature...
OPERATING INSTRUCTIONS Defrost cycles Refrigerator There is no need for routine defrosting of the unit as this occurs automatically as follows: 1. Cycle defrost To keep the temperature stable inside the chamber, the refrigeration compressor is cycled on and off. During “off”...
TEMPERATURE RECORDER Automatic temperature recorder To record the chamber temperature, an optional automatic temperature recorder is available. Please consult with our sales department or agency for recorder installation. For the proper usage of temperature recorder, refer to an instruction manual included with the recorder. The available recorder and a dedicated fixture is as follows: Temperature recorder: Model MTR-0620LH (for refrigerator)
Page 18
TEMPERATURE RECORDER Attachment of temperature recorder For refrigerator Fig. 1 1. Attach the fixture to the temperature recorder with two screws. (Fig. 1) Fig. 2 2. Remove the screws at the right corner on the top of the unit and fix the recorder to the position. (Fig. 2) 3.
REMOTE ALARM TERMINAL WARING Always disconnect the power supply to the unit before connection of alarm equipment to remote alarm terminals to prevent electric shock or injury. The terminal of the remote alarm is installed in the terminal cover at rear of the cabinet. The maximum contact rating is DC 30 V, 2 A.
TROUBLE SHOOTING If the unit malfunctions, check out the following before calling for service. If nothing operates even when plug in 1. The unit is not connected to the power supply or capacity of power source enough. 2. There is a power failure. 3.
250V, 15W (AC 220 to 240V) Accessories Key 1set, Nylon clip; large 2 pcs., small 2 pcs. Weight MPR-411F; 124 Kg, MPR-411FR; 135 Kg Option Automatic temperature recorder (for refrigerator MTR-0620LH, for freezer MTR-4014LH) Mounting kit for temperature recorder [MDF-53RO(H)]...
PERFORMANCE Refrigerator Freezer Control range C to 14 C to -30 (Ambient temp.; -5 C to +35 C, No load) (Ambient temp.; -5 C to +35 C, No load) Noise level 39 dB (A scale) Maximum pressure 1705 kPa 1901 kPa Rated voltage AC 110 V AC 115 V...
Procedure to be adhered to in order to reduce safety risk indicated in b) below. Date : Signature : Address, Division : Telephone : Product name : Model : Serial number : Date of Installation : Pharmaceutical MPR-411F Refrigerator with Freezer MPR-411FR Please decontaminate the unit yourself before calling the service engineer.
Page 24